To Use Reading Hold
The reading hold feature allows you to capture and hold a stable
reading on the display. When a stable reading is detected, the
multimeter emits a beep and holds the value on the display.
•
Reading hold has an adjustable
sensitivity band
to allow you to
select which readings are considered stable enough to be displayed.
The band is expressed as a percent of reading on the selected range.
The multimeter will capture and display a new value only after
three
consecutive readings are within the band.
•
The default band is 0.10% of reading. After enabling reading hold,
you can choose a different band by pressing
Shift >
(Menu Recall). This takes you to the “
READ HOLD
” command in
the
TRIG MENU
(
only if reading hold is enabled
).
Go down to the “parameter” level, and then select a value:
0.01%,
0.10%
, 1.00%, or 10.00% of reading.
•
The sensitivity band is stored in
volatile
memory; the multimeter
sets the band to 0.10% of reading when power has been off or after a
remote interface reset.
Toggles between auto trigger
and reading hold.
Hold annunciator is on when
reading hold is enabled.
To adjust the reading hold sensitivity band,
use the TRIG menu.
2
Chapter 2 Front-Panel Menu Operation
To Use Reading Hold
43
Summary of Contents for 34401A
Page 12: ...1 Quick Start 1 ...
Page 26: ...2 Front Panel Menu Operation 2 ...
Page 50: ...3 Features and Functions 3 ...
Page 103: ...4 Remote Interface Reference 4 ...
Page 135: ...SCPI Status System 4 Chapter 4 Remote Interface Reference The SCPI Status Model 135 ...
Page 170: ...5 Error Messages 5 ...
Page 182: ...6 Application Programs 6 ...
Page 196: ...7 Measurement Tutorial 7 ...
Page 214: ...8 Specifications 8 ...