Measurement Configuration
This section contains information to help you configure the multimeter
for making measurements. You may never have to change any of the
measurement parameters discussed here, but they are provided to give
you the flexibility you might need.
AC Signal Filter
The multimeter uses three different ac filters which enable you to either
optimize low frequency accuracy or achieve faster ac settling times.
The multimeter selects the
slow
,
medium
, or
fast
filter based on the
input frequency that you specify.
Applies to ac voltage and ac current measurements only.
•
The ac filter selection is stored in
volatile
memory; the multimeter
selects the
medium
filter (20 Hz) when power has been off or after a
remote interface reset.
•
Front-Panel Operation
: Select from the menu the slow filter (3 Hz),
medium filter (
20 Hz
), or fast filter (200 Hz).
The default is the
medium filter.
1: AC FILTER (MEAS MENU)
•
Remote Interface Operation
: Specify the lowest frequency expected in
the input signal. The multimeter selects the appropriate filter based
on the frequency you specify (
see table above
). The
CONFigure
and
MEASure?
commands select the 20 Hz filter.
DETector:BANDwidth {3|20|200|MINimum|MAXimum}
Input Frequency
3 Hz to 300 kHz
20 Hz to 300 kHz
200 Hz to 300 kHz
AC Filter Selected
Slow filter
Medium filter (default)
Fast filter
Settling Time
7 seconds / reading
1 reading / second
10 readings / second
3
Chapter 3 Features and Functions
Measurement Configuration
51
Summary of Contents for 34401A
Page 12: ...1 Quick Start 1 ...
Page 26: ...2 Front Panel Menu Operation 2 ...
Page 50: ...3 Features and Functions 3 ...
Page 103: ...4 Remote Interface Reference 4 ...
Page 135: ...SCPI Status System 4 Chapter 4 Remote Interface Reference The SCPI Status Model 135 ...
Page 170: ...5 Error Messages 5 ...
Page 182: ...6 Application Programs 6 ...
Page 196: ...7 Measurement Tutorial 7 ...
Page 214: ...8 Specifications 8 ...