Remote Interface Operation
: You can use the following commands to
make min-max measurements.
CALCulate:FUNCtion AVERage
CALCulate:STATe {OFF|ON}
CALCulate:AVERage:MINimum?
read the minimum value
CALCulate:AVERage:MAXimum?
read the maximum value
CALCulate:AVERage:AVERage?
read the average of all readings
CALCulate:AVERage:COUNt?
read the count
A new command is available starting with firmware Revision 2 which
allows you to take readings using
INITiate
without storing them in
internal memory. This command may be useful with the min-max
operation since it allows you to determine the average of a series of
readings without storing the individual values.
DATA:FEED RDG_STORE, ""
do not store readings
DATA:FEED RDG_STORE, "CALCulate"
store readings (default)
See page 126 for more information on using the
DATA:FEED
command.
Null (Relative) Operation
When making null measurements, also called
relative
, each reading
is the difference between a stored null value and the input signal.
One possible application is in making more accurate two-wire ohms
measurements by nulling the test lead resistance.
Result
=
reading
–
null value
Applies to all measurement functions, except continuity, diode, and ratio.
•
The null value is adjustable and you can set it to any value between
0 and
±
120% of the highest range, for the present function.
•
The null value is stored in
volatile
memory; the value is cleared
when power has been off, after a remote interface reset, or after a
function change.
3
Chapter 3 Features and Functions
Math Operations
65
Summary of Contents for 34401A
Page 12: ...1 Quick Start 1 ...
Page 26: ...2 Front Panel Menu Operation 2 ...
Page 50: ...3 Features and Functions 3 ...
Page 103: ...4 Remote Interface Reference 4 ...
Page 135: ...SCPI Status System 4 Chapter 4 Remote Interface Reference The SCPI Status Model 135 ...
Page 170: ...5 Error Messages 5 ...
Page 182: ...6 Application Programs 6 ...
Page 196: ...7 Measurement Tutorial 7 ...
Page 214: ...8 Specifications 8 ...