Voltmeter Complete Terminal
The rear-panel
VM Comp
(voltmeter complete) terminal provides a
low-true pulse after the completion of each measurement. Voltmeter
complete and external trigger (
see below
) implement a standard hardware
handshake sequence between measurement and switching devices.
External Trigger Terminal
You can trigger the multimeter by applying a low-true pulse to the
rear-panel
Ext Trig
(external trigger) terminal. To use this terminal
from the remote interface, you must select the external trigger source
(
TRIGger:SOURce EXTernal
).
You can use a simple switch to generate an external trigger using the
Ext Trig
input as shown below.
5 V
0 V
Approximately
2
µ
s
Output
5 V
0 V
>1
µ
s
Input
Ext Trig
3
Chapter 3 Features and Functions
Triggering
83