Limit Testing
The limit test operation enables you to perform pass/fail testing to
upper and lower limits that you specify.
Applies to all measurement functions, except continuity and diode tests.
•
You can set the upper and lower limits to any value between 0 and
±
120% of the highest range, for the present function. The upper limit
selected should always be a more positive number than the lower
limit.
The default upper and lower limits are both “0”.
•
The upper and lower limits are stored in
volatile
memory; the
multimeter sets both limits to “0” when power has been off, after a
remote interface reset, or after a function change.
•
You can configure the multimeter to generate a service request (
SRQ
)
on the first occurrence of a failed reading.
See “The SCPI Status
Model,” starting on page 134 for more information.
•
Front-Panel Operation
: The multimeter displays “
OK
” for each
reading that is within the specified limits. It displays “
HI
” or “
LO
” for
each reading that exceeds the upper or lower limit. The multimeter
beeps once (if the front-panel beeper is enabled) on the first
occurrence of a failed reading after a good reading.
See also “Beeper
Control,” on page 88.
5: LIMIT TEST (MATH MENU)
enable or disable limit test
6: HIGH LIMIT (MATH MENU)
set the upper limit
7: LOW LIMIT (MATH MENU)
set the lower limit
You can also turn off limit test by selecting a different math operation
from the front panel (only one math operation can be enabled at a time).
3
Chapter 3 Features and Functions
Math Operations
69
Summary of Contents for 34401A
Page 12: ...1 Quick Start 1 ...
Page 26: ...2 Front Panel Menu Operation 2 ...
Page 50: ...3 Features and Functions 3 ...
Page 103: ...4 Remote Interface Reference 4 ...
Page 135: ...SCPI Status System 4 Chapter 4 Remote Interface Reference The SCPI Status Model 135 ...
Page 170: ...5 Error Messages 5 ...
Page 182: ...6 Application Programs 6 ...
Page 196: ...7 Measurement Tutorial 7 ...
Page 214: ...8 Specifications 8 ...