The multimeter holds the
DTR
line
FALSE
while output is suspended.
A form of interface
deadlock
exists until the controller asserts the
DSR
line
TRUE
to allow the multimeter to complete the transmission.
You can break the interface deadlock by sending the <
Ctrl-C
> character,
which clears the operation in progress and discards pending output
(this is equivalent to the
IEEE
-488 device clear action).
For the <Ctrl-C>
character to be recognized reliably by the multimeter while it holds DTR
FALSE, the controller must first set DSR FALSE.
In addition, you may have difficulty sending the <
Ctrl-C
> character if
you are interrupting a query operation, in which case the multimeter
hold the
DTR
line
FALSE
. This may prevent the controller from sending
anything unless you first reprogram the interface to ignore
DTR
.
RS-232 Troubleshooting
Here are a few things to check if you are having problems communicating
over the
RS
-232 interface. If you need additional help, refer to the
documentation that came with your computer.
•
Verify that the multimeter and your computer are configured for the
same baud rate, parity, and number of data bits. Make sure that your
computer is set up for
1 start bit
and
2 stop bits
(these values are fixed
on the multimeter).
•
Make sure to execute the
SYSTem:REMote
command to place the
multimeter in the
REMOTE
mode.
•
Verify that you have connected the correct interface cable and adapters.
Even if the cable has the proper connectors for your system, the
internal wiring may not be correct. The
HP 34398A Cable Kit
can be
used to connect the multimeter to most computers or terminals.
•
Verify that you have connected the interface cable to the correct
serial port on your computer (
COM
1,
COM
2, etc).
Chapter 4 Remote Interface Reference
RS-232 Interface Configuration
152
Summary of Contents for 34401A
Page 12: ...1 Quick Start 1 ...
Page 26: ...2 Front Panel Menu Operation 2 ...
Page 50: ...3 Features and Functions 3 ...
Page 103: ...4 Remote Interface Reference 4 ...
Page 135: ...SCPI Status System 4 Chapter 4 Remote Interface Reference The SCPI Status Model 135 ...
Page 170: ...5 Error Messages 5 ...
Page 182: ...6 Application Programs 6 ...
Page 196: ...7 Measurement Tutorial 7 ...
Page 214: ...8 Specifications 8 ...