Execution Errors
-101
Invalid character
An invalid character was found in the command string. You may have
inserted a character such as #, $, or % in the command header or within
a parameter. Example:
CONF:VOLT#DC
-102
Syntax error
Invalid syntax was found in the command string. You may have
inserted a blank space before or after a colon in the command header,
or before a comma. Example:
SAMP:COUN ,1
-103
Invalid separator
An invalid separator was found in the command string. You may have
used a comma instead of a colon, semicolon, or blank space – or you may
have used a blank space instead of a comma. Example:
TRIG:COUN,1
or
CONF:FREQ 1000 0.1
-104
Data type error
The wrong parameter type was found in the command string. You may
have specified a number where a string was expected, or vice versa.
Example:
DISP:TEXT 5.0
-105
GET not allowed
A Group Execute Trigger (
GET
) is not allowed within a command string.
-108
Parameter not allowed
More parameters were received than expected for the command.
You may have entered an extra parameter, or you added a parameter
to a command that does not accept a parameter. Example:
READ? 10
-109
Missing parameter
Fewer parameters were received than expected for the command.
You omitted one or more parameters that are required for this
command. Example:
SAMP:COUN
5
Chapter 5 Error Messages
Execution Errors
173
Summary of Contents for 34401A
Page 12: ...1 Quick Start 1 ...
Page 26: ...2 Front Panel Menu Operation 2 ...
Page 50: ...3 Features and Functions 3 ...
Page 103: ...4 Remote Interface Reference 4 ...
Page 135: ...SCPI Status System 4 Chapter 4 Remote Interface Reference The SCPI Status Model 135 ...
Page 170: ...5 Error Messages 5 ...
Page 182: ...6 Application Programs 6 ...
Page 196: ...7 Measurement Tutorial 7 ...
Page 214: ...8 Specifications 8 ...