Power-On and Reset State
Measurement Configuration
AC
Filter
Autozero
•
Continuity Threshold
Function
Input
Resistance
Integration Time
Range
Resolution
Math Operations
Math State, Function
Math Registers
•
dBm Reference Resistance
Triggering Operations
Reading Hold Threshold
Samples Per Trigger
Trigger Delay
Trigger Source
System-Related Operations
•
Beeper Mode
•
Comma Separators
Display Mode
Reading
Memory
Input/Output Configuration
•
Baud Rate
•
HP-IB Address
•
Interface
•
Language
•
Parity
Calibration
•
Calibration State
Power-On/Reset State
20 Hz (medium filter)
On
•
10
Ω
DC
volts
10
M
Ω
(fixed for all dcv ranges)
10
PLCs
Autorange
5
1
⁄
2
digits, slow mode
Power-On/Reset State
Off, Null
All registers are cleared
•
600
Ω
Power-On/Reset State
0.10% of range
1
sample
Automatic Delay
Auto
Trigger
Power-On/Reset State
•
On
•
On
On
Off (cleared)
Power-On/Reset State
•
9600 baud
•
22
•
HP-IB (IEEE-488)
•
SCPI
•
Even (7 data bits)
Power-On/Reset State
•
Secured
For your convenience,
this table is duplicated
on the rear cover of this
manual and on the
Quick Reference Card.
The parameters marked with a bullet (
•
) are stored in
non-volatile
memory.
The factory settings are shown.
3
Chapter 3 Features and Functions
Power-On and Reset State
101
Summary of Contents for 34401A
Page 12: ...1 Quick Start 1 ...
Page 26: ...2 Front Panel Menu Operation 2 ...
Page 50: ...3 Features and Functions 3 ...
Page 103: ...4 Remote Interface Reference 4 ...
Page 135: ...SCPI Status System 4 Chapter 4 Remote Interface Reference The SCPI Status Model 135 ...
Page 170: ...5 Error Messages 5 ...
Page 182: ...6 Application Programs 6 ...
Page 196: ...7 Measurement Tutorial 7 ...
Page 214: ...8 Specifications 8 ...