CONFigure:VOLTage:DC {<
range
>|MIN|MAX|DEF},{<
resolution
>|MIN|MAX|DEF}
Preset and configure the multimeter for dc voltage measurements with
the specified range and resolution. This command
does not
initiate the
measurement.
CONFigure:VOLTage:DC:RATio {<
range
>|MIN|MAX|DEF},{<
resolution
>|MIN|MAX|DEF}
Preset and configure the multimeter for dc:dc ratio measurements with
the specified range and resolution. This command
does not
initiate the
measurement. For ratio measurements, the specified range applies to
the signal connected to the
Input
terminals. Autoranging is automatically
selected for reference voltage measurements on the
Sense
terminals.
CONFigure:VOLTage:AC {<
range
>|MIN|MAX|DEF},{<
resolution
>|MIN|MAX|DEF}
Preset and configure the multimeter for ac voltage measurements with
the specified range and resolution. This command
does not
initiate the
measurement. For ac measurements, resolution is actually fixed at
6
1
⁄
2
digits. The
resolution
parameter only affects the front-panel display.
CONFigure:CURRent:DC {<
range
>
|MIN|MAX|DEF},{<
resolution
>|MIN|MAX|DEF}
Preset and configure the multimeter for dc current measurements with
the specified range and resolution. This command
does not
initiate the
measurement.
CONFigure:CURRent:AC {<
range
>|MIN|MAX|DEF},{<
resolution
>|MIN|MAX|DEF}
Preset and configure the multimeter for ac current measurements with
the specified range and resolution. This command
does not
initiate the
measurement. For ac measurements, resolution is actually fixed at
6
1
⁄
2
digits. The
resolution
parameter only affects the front-panel display.
CONFigure:RESistance {<
range
>|MIN|MAX|DEF},{<
resolution
>|MIN|MAX|DEF}
Preset and configure the multimeter for 2-wire ohms measurements
with the specified range and resolution. This command
does not
initiate
the measurement.
CONFigure:FRESistance {<
range
>|MIN|MAX|DEF},{<
resolution
>|MIN|MAX|DEF}
Preset and configure the multimeter for 4-wire ohms measurements
with the specified range and resolution. This command
does not
initiate
the measurement.
4
Chapter 4 Remote Interface Reference
The MEASure? and CONFigure Commands
119
Summary of Contents for 34401A
Page 12: ...1 Quick Start 1 ...
Page 26: ...2 Front Panel Menu Operation 2 ...
Page 50: ...3 Features and Functions 3 ...
Page 103: ...4 Remote Interface Reference 4 ...
Page 135: ...SCPI Status System 4 Chapter 4 Remote Interface Reference The SCPI Status Model 135 ...
Page 170: ...5 Error Messages 5 ...
Page 182: ...6 Application Programs 6 ...
Page 196: ...7 Measurement Tutorial 7 ...
Page 214: ...8 Specifications 8 ...