Remote Interface Configuration
This section gives information on configuring the remote interface.
For additional information, see chapter 4, “Remote Interface Reference,”
starting on page 103.
HP-IB Address
Each device on the
HP-IB
(
IEEE-488
) interface must have a unique
address. You can set the multimeter’s address to any value between
0 and 31. The address is set to “
22
” when the multimeter is shipped
from the factory. The
HP-IB
address is displayed at power-on.
The HP-IB address can be set only from the front-panel.
•
The address is stored in
non-volatile
memory, and
does not
change
when power has been off or after a remote interface reset.
•
You can set the address to “31” which is the
talk only
mode. In this
mode, the multimeter can output readings directly to a printer
without being addressed by a bus controller (over either
HP-IB
or
RS-232
). For proper operation, make sure your printer is configured in
the
listen always
mode. Address 31 is not a valid address if you are
operating the multimeter from the
HP-IB
interface with a bus controller.
If you select the
RS-232
interface and then set the
HP-IB
address to the
talk only address (31), the multimeter will
send
readings over the
RS-232
interface when in the local mode.
•
If you select the
RS-232
interface and then set the
HP-IB
address to
the talk only address (31), the multimeter will
send
readings over the
RS-232
interface when in the local mode.
•
Your
HP-IB
bus controller has its own address. Be sure to avoid using
the bus controller’s address for
any
instrument on the interface bus.
Hewlett-Packard controllers generally use address “21”.
•
Front-Panel Operation
:
1: HP-IB ADDR (I/O MENU)
See also “To Set the
HP-IB
Address,” on page 161.
3
Chapter 3 Features and Functions
Remote Interface Configuration
91
Summary of Contents for 34401A
Page 12: ...1 Quick Start 1 ...
Page 26: ...2 Front Panel Menu Operation 2 ...
Page 50: ...3 Features and Functions 3 ...
Page 103: ...4 Remote Interface Reference 4 ...
Page 135: ...SCPI Status System 4 Chapter 4 Remote Interface Reference The SCPI Status Model 135 ...
Page 170: ...5 Error Messages 5 ...
Page 182: ...6 Application Programs 6 ...
Page 196: ...7 Measurement Tutorial 7 ...
Page 214: ...8 Specifications 8 ...