CALCulate:AVERage:MINimum?
Read the minimum value found during a min-max operation. The
multimeter clears the value when min-max is turned on, when power
has been off, or after a remote interface reset.
[Stored in volatile memory]
CALCulate:AVERage:MAXimum?
Read the maximum value found during a min-max operation. The
multimeter clears the value when min-max is turned on, when power
has been off, or after a remote interface reset.
[Stored in volatile memory]
CALCulate:AVERage:AVERage?
Read the average of all readings taken since min-max was enabled.
The multimeter clears the value when min-max is turned on, when
power has been off, or after a remote interface reset.
[Stored in volatile
memory]
CALCulate:AVERage:COUNt?
Read the number of readings taken since min-max was enabled. The
multimeter clears the value when min-max is turned on, when power
has been off, or after a remote interface reset.
[Stored in volatile
memory]
CALCulate:NULL:OFFSet {<
value
>|MINimum|MAXimum}
Store a null value in the multimeter’s Null Register.
You must turn on the math
operation before writing to the math register
. You can set the null value to any
number between 0 and
±
120% of the highest range, for the present
function.
MIN
= –120% of the highest range.
MAX
= 120% of the highest
range.
[Stored in volatile memory]
CALCulate:NULL:OFFSet? [MINimum|MAXimum]
Query the null value.
CALCulate:DB:REFerence {<
value
>|MINimum|MAXimum}
Store a relative value in the dB Relative Register.
You must turn on the math
operation before writing to the math register
. You can set the relative value to
any number between 0 dBm and
±
200 dBm.
MIN
= –200.00 dBm.
MAX
= 200.00 dBm.
[Stored in volatile memory]
CALCulate:DB:REFerence? [MINimum|MAXimum]
Query the dB relative value.
4
Chapter 4 Remote Interface Reference
Math Operation Commands
125
Summary of Contents for 34401A
Page 12: ...1 Quick Start 1 ...
Page 26: ...2 Front Panel Menu Operation 2 ...
Page 50: ...3 Features and Functions 3 ...
Page 103: ...4 Remote Interface Reference 4 ...
Page 135: ...SCPI Status System 4 Chapter 4 Remote Interface Reference The SCPI Status Model 135 ...
Page 170: ...5 Error Messages 5 ...
Page 182: ...6 Application Programs 6 ...
Page 196: ...7 Measurement Tutorial 7 ...
Page 214: ...8 Specifications 8 ...