The MEASure? and CONFigure Commands
See also “Measurement Configuration,” starting on page 51 in chapter 3.
•
For the
range
parameter,
MIN
selects the lowest range for the
selected function;
MAX
selects the highest range;
DEF
selects
autoranging.
•
For the
resolution
parameter, specify the resolution in the same units
as the measurement function,
not in number of digits
.
MIN
selects the
smallest value accepted, which gives the best resolution;
MAX
selects
the largest value accepted, which gives the least resolution;
DEF
selects the default resolution which is 5
1
⁄
2
digits slow (10
PLC
).
Note
: You must specify a
range
to use the
resolution
parameter.
MEASure:VOLTage:DC? {<
range
>|
MIN|MAX|DEF},{<
resolution
>|MIN|MAX|DEF}
Preset and make a dc voltage measurement with the specified range
and resolution. The reading is sent to the output buffer.
MEASure:VOLTage:DC:RATio? {<
range
>|MIN|MAX|DEF},{<
resolution
>|MIN|MAX|DEF}
Preset and make a dc:dc ratio measurement with the specified range
and resolution. The reading is sent to the output buffer. For ratio
measurements, the specified range applies to the signal connected to the
Input
terminals. Autoranging is automatically selected for reference
voltage measurements on the
Sense
terminals.
MEASure:VOLTage:AC? {<
range
>|MIN|MAX|DEF},{<
resolution
>|MIN|MAX|DEF}
Preset and make an ac voltage measurement with the specified range
and resolution. The reading is sent to the output buffer. For ac
measurements, resolution is actually fixed at 6
1
⁄
2
digits. The
resolution
parameter only affects the front-panel display.
MEASure:CURRent:DC? {<
range
>|MIN|MAX|DEF},{<
resolution
>|MIN|MAX|DEF}
Preset and make a dc current measurement with the specified range
and resolution. The reading is sent to the output buffer.
4
Chapter 4 Remote Interface Reference
The MEASure? and CONFigure Commands
117
Summary of Contents for 34401A
Page 12: ...1 Quick Start 1 ...
Page 26: ...2 Front Panel Menu Operation 2 ...
Page 50: ...3 Features and Functions 3 ...
Page 103: ...4 Remote Interface Reference 4 ...
Page 135: ...SCPI Status System 4 Chapter 4 Remote Interface Reference The SCPI Status Model 135 ...
Page 170: ...5 Error Messages 5 ...
Page 182: ...6 Application Programs 6 ...
Page 196: ...7 Measurement Tutorial 7 ...
Page 214: ...8 Specifications 8 ...