SCPI Language Version Query
The multimeter complies with the rules and regulations of the present
version of
SCPI
(Standard Commands for Programmable Instruments).
You can determine the
SCPI
version with which the multimeter is in
compliance by sending a command from the remote interface.
You cannot query the
SCPI
version from the front panel.
•
The following command returns the
SCPI
version.
SYSTem:VERSion?
Returns a string in the form “YYYY.V” where the “Y’s” represent the
year of the version, and the “V” represents a version number for that
year (for example, 1991.0).