CONFigure:FREQuency {<
range
>|MIN|MAX|DEF},{<
resolution
>|MIN|MAX|DEF}
Preset and configure a frequency measurement with the specified range
and resolution. This command
does not
initiate the measurement.
For frequency measurements, the multimeter uses one “range” for all
inputs between 3 Hz and 300 kHz. With no input signal applied,
frequency measurements return “0”.
CONFigure:PERiod {<
range
>|MIN|MAX|DEF},{<
resolution
>|MIN|MAX|DEF}
Preset and configure a period measurement with the specified range
and resolution. This command
does not
initiate the measurement.
For period measurements, the multimeter uses one “range” for all
inputs between 0.33 seconds and 3.3
µ
sec. With no input signal applied,
period measurements return “0”.
CONFigure:CONTinuity
Preset and configure the multimeter for continuity measurements.
This command
does not
initiate the measurement. The range and
resolution are fixed for continuity tests (1 k
Ω
range and 5
1
⁄
2
digits).
CONFigure:DIODe
Preset and configure the multimeter for diode measurements.
This command
does not
initiate the measurement. The range and
resolution are fixed for diode tests (1 Vdc range with 1 mA current
source output and 5
1
⁄
2
digits).
CONFigure?
Query the multimeter’s present configuration and return a quoted string.
Chapter 4 Remote Interface Reference
The MEASure? and CONFigure Commands
120
Summary of Contents for 34401A
Page 12: ...1 Quick Start 1 ...
Page 26: ...2 Front Panel Menu Operation 2 ...
Page 50: ...3 Features and Functions 3 ...
Page 103: ...4 Remote Interface Reference 4 ...
Page 135: ...SCPI Status System 4 Chapter 4 Remote Interface Reference The SCPI Status Model 135 ...
Page 170: ...5 Error Messages 5 ...
Page 182: ...6 Application Programs 6 ...
Page 196: ...7 Measurement Tutorial 7 ...
Page 214: ...8 Specifications 8 ...