-221
Settings conflict
This error can be generated in one of the following situations:
•
You sent a
CONFigure
or
MEASure
command with autorange enabled
and with a fixed resolution. Example:
CONF:VOLT:DC DEF,0.1
•
You turned math on (
CALC:STAT ON
) and then changed to a math
operation that was not valid with the present measurement function.
For example, dB measurements are not allowed with 2-wire ohms.
The math state is turned off as a result of this condition.
-222
Data out of range
A numeric parameter value is outside the valid range for the command.
Example:
TRIG:COUN -3
-223
Too much data
A character string was received but could not be executed because the
string length was more than 12 characters. This error can be generated
by the
CALibration:STRing
and
DISPlay:TEXT
commands.
-224
Illegal parameter value
A discrete parameter was received which was not a valid choice for
the command. You may have used an invalid parameter choice.
Example:
CALC:FUNC SCALE
(
SCALE
is not a valid choice).
-230
Data stale
A
FETCh?
command was received but internal reading memory was
empty. The reading retrieved may be invalid.
-330
Self-test failed
The multimeter’s complete self-test failed from the remote interface
(
*TST?
command). In addition to this error, more specific self-test errors
are also reported.
See also “Self-Test Errors,” starting on page 179.
Chapter 5 Error Messages
Execution Errors
176
Summary of Contents for 34401A
Page 12: ...1 Quick Start 1 ...
Page 26: ...2 Front Panel Menu Operation 2 ...
Page 50: ...3 Features and Functions 3 ...
Page 103: ...4 Remote Interface Reference 4 ...
Page 135: ...SCPI Status System 4 Chapter 4 Remote Interface Reference The SCPI Status Model 135 ...
Page 170: ...5 Error Messages 5 ...
Page 182: ...6 Application Programs 6 ...
Page 196: ...7 Measurement Tutorial 7 ...
Page 214: ...8 Specifications 8 ...