To Make Null (Relative) Measurements
Each null measurement, also called
relative,
is the difference between a
stored null value and the input signal.
Result
= reading
–
null value
•
You can make null measurements with any function except
continuity, diode, or ratio. The null operation is local to the selected
function; when you change functions, null is disabled.
•
To null the test lead resistance for more accurate two-wire ohms
measurements, short the ends of the test leads together and then
press
Null
.
•
The first reading taken after you press
Null
is stored as the null
value in the Null Register. Any previously stored value is
replaced with the new value.
•
After enabling null, you can edit the stored null value by
pressing
Shift >
(Menu Recall). This takes you to the
“
NULL VALUE
” command in the
MATH MENU
(
only if null is
enabled
). Go down to the “parameter” level, and then edit the
displayed value.
•
The null register is cleared when you change functions, turn null off,
turn off the power, or perform a remote interface reset.
Enables null operation;
Press again to disable.
Math annunciator is on when
null operation is enabled.
To read / edit the null value, use the MATH menu.
Chapter 2 Front-Panel Menu Operation
To Make Null (Relative) Measurements
38
Summary of Contents for 34401A
Page 12: ...1 Quick Start 1 ...
Page 26: ...2 Front Panel Menu Operation 2 ...
Page 50: ...3 Features and Functions 3 ...
Page 103: ...4 Remote Interface Reference 4 ...
Page 135: ...SCPI Status System 4 Chapter 4 Remote Interface Reference The SCPI Status Model 135 ...
Page 170: ...5 Error Messages 5 ...
Page 182: ...6 Application Programs 6 ...
Page 196: ...7 Measurement Tutorial 7 ...
Page 214: ...8 Specifications 8 ...