Trigger Delay
You can insert a delay between the trigger signal and each sample that
follows. This may be useful in applications where you want to allow the
input to settle before taking a reading, or for pacing a burst of readings.
If you do not specify a trigger delay, the multimeter automatically
selects a delay for you.
•
Delay range: 0 to 3600 seconds.
The default trigger delay is
automatic; the delay is determined by function, range, integration time,
and ac filter setting (see also “Automatic Trigger Delays,” on page 81).
•
The trigger delay is stored in
volatile
memory; the multimeter selects
an automatic trigger delay when power has been off or after a remote
interface reset. The
CONFigure
and
MEASure?
commands
automatically set the trigger delay to automatic.
•
If you specify a delay other than automatic,
that same delay
is used
for
all
functions and ranges.
•
If you have configured the multimeter to take more than one reading
per trigger (sample count > 1), the specified trigger delay is inserted
between the trigger and
each
reading.
•
Front-Panel Operation
: You can use an automatic trigger delay or
you can specify a delay in seconds.
2: TRIG DELAY (TRIG MENU)
If an automatic trigger delay is enabled, “
AUTO
” is displayed
momentarily before the actual number of seconds is displayed.
--- AUTO ---
3
Chapter 3 Features and Functions
Triggering
79
Summary of Contents for 34401A
Page 12: ...1 Quick Start 1 ...
Page 26: ...2 Front Panel Menu Operation 2 ...
Page 50: ...3 Features and Functions 3 ...
Page 103: ...4 Remote Interface Reference 4 ...
Page 135: ...SCPI Status System 4 Chapter 4 Remote Interface Reference The SCPI Status Model 135 ...
Page 170: ...5 Error Messages 5 ...
Page 182: ...6 Application Programs 6 ...
Page 196: ...7 Measurement Tutorial 7 ...
Page 214: ...8 Specifications 8 ...