10.216,5 VDC
This is the 10 Vdc range, 5
1
⁄
2
digits are displayed.
-
045.23 mVDC
This is the 100 mVdc range, 4
1
⁄
2
digits are displayed.
113.325,6 OHM
This is the 100 ohm range, 6
1
⁄
2
digits are displayed.
•
The resolution is stored in
volatile
memory; the multimeter sets the
resolution to 5
1
⁄
2
digits (for all functions) when power has been off or
after a remote interface reset.
•
The resolution is fixed at 5
1
⁄
2
digits for continuity and diode tests.
•
For dc and resistance measurements, changing the number of digits
does more than just change the resolution of the multimeter. It also
changes the
integration time,
which is the period the multimeter’s
analog-to-digital (A/D) converter samples the input signal for a
measurement.
See also “Integration Time,” on page 57.
•
For ac measurements, the resolution is actually fixed at 6
1
⁄
2
digits.
If you select 4
1
⁄
2
digits or 5
1
⁄
2
digits, the multimeter “masks” one or
two digits. The only way to control the reading rate for ac measurements
is by setting a trigger delay (
see page 79
).
•
For ratio measurements, the specified resolution applies to the signal
connected to the
Input
terminals.
“
1
⁄
2
” digit
5 digits
“
1
⁄
2
” digit
3
Chapter 3 Features and Functions
Measurement Configuration
55
Summary of Contents for 34401A
Page 12: ...1 Quick Start 1 ...
Page 26: ...2 Front Panel Menu Operation 2 ...
Page 50: ...3 Features and Functions 3 ...
Page 103: ...4 Remote Interface Reference 4 ...
Page 135: ...SCPI Status System 4 Chapter 4 Remote Interface Reference The SCPI Status Model 135 ...
Page 170: ...5 Error Messages 5 ...
Page 182: ...6 Application Programs 6 ...
Page 196: ...7 Measurement Tutorial 7 ...
Page 214: ...8 Specifications 8 ...