*ESR?
Query the Standard event register. The multimeter returns a decimal
value which corresponds to the binary-weighted sum of all bits set in the
register.
*OPC
Sets the “operation complete” bit (bit 0) in the Standard Event register
after the command is executed.
*OPC?
Returns “1” to the output buffer after the command is executed.
*PSC {0|1}
Power-on status clear. Clear the Status Byte and Standard Event
register enable masks when power is turned on (
*PSC 1
). When
*PSC 0
is in effect, the Status Byte and Standard Event register enable masks
are not
cleared when power is turned on.
[Stored in non-volatile memory]
*PSC?
Query the power-on status clear setting. Returns “0” (
*PSC 0
) or
“1” (
*PSC 1
).
*SRE <
enable value
>
Enable bits in the Status Byte enable register.
*SRE?
Query the Status Byte enable register. The multimeter returns a
decimal value which corresponds to the binary-weighted sum of all bits
set in the register.
*STB?
Query the Status Byte summary register. The
*STB?
command is
similar to a serial poll but it is processed like any other instrument
command. The
*STB?
command returns the same result as a serial
poll but the “request service” bit (bit 6)
is not
cleared if a serial poll
has occurred.
4
Chapter 4 Remote Interface Reference
Status Reporting Commands
145
Summary of Contents for 34401A
Page 12: ...1 Quick Start 1 ...
Page 26: ...2 Front Panel Menu Operation 2 ...
Page 50: ...3 Features and Functions 3 ...
Page 103: ...4 Remote Interface Reference 4 ...
Page 135: ...SCPI Status System 4 Chapter 4 Remote Interface Reference The SCPI Status Model 135 ...
Page 170: ...5 Error Messages 5 ...
Page 182: ...6 Application Programs 6 ...
Page 196: ...7 Measurement Tutorial 7 ...
Page 214: ...8 Specifications 8 ...