Internal Triggering
In the internal trigger mode (remote interface only),
the trigger signal is always present. When you place the multimeter in
the wait-for-trigger state, the trigger is issued immediately.
This is the
power-on trigger source for remote interface operation.
To select the internal trigger source, send the following command.
The
CONFigure
and
MEASure?
commands automatically set the
trigger source to
IMMediate
.
TRIGger:SOURce IMMediate
Software (Bus) Triggering
The bus trigger mode is available only
from the remote interface. This mode is similar to the single trigger
mode from the front panel, but you trigger the multimeter by sending
a bus trigger command.
•
To select the bus trigger source, send the following command.
TRIGger:SOURce BUS
•
To trigger the multimeter from the remote interface (
HP-IB
or
RS-232
),
send the
*TRG
(trigger) command. The
*TRG
command
will not
be
accepted unless the multimeter is in the wait-for-trigger state.
•
You can also trigger the multimeter from the
HP-IB
interface by
sending the
IEEE-488
Group Execute Trigger (
GET
) message.
The multimeter must be in the wait-for-trigger state. The following
statement shows how to send a
GET
using
HP BASIC
.
TRIGGER 722
Group Execute Trigger
3
Chapter 3 Features and Functions
Triggering
75
Summary of Contents for 34401A
Page 12: ...1 Quick Start 1 ...
Page 26: ...2 Front Panel Menu Operation 2 ...
Page 50: ...3 Features and Functions 3 ...
Page 103: ...4 Remote Interface Reference 4 ...
Page 135: ...SCPI Status System 4 Chapter 4 Remote Interface Reference The SCPI Status Model 135 ...
Page 170: ...5 Error Messages 5 ...
Page 182: ...6 Application Programs 6 ...
Page 196: ...7 Measurement Tutorial 7 ...
Page 214: ...8 Specifications 8 ...