9. The   internal   wiring   of   the  unit  should  not  be  

tampered with or altered.  Doing so will likely         

the unit and increase risk of electric shock.

10. Replace the battery as soon as the battery 

indicator “       ” appears. When working under a low 

battery, the unit might produce false readings that can 

lead to electric shock or personal injury.

11. Turn the unit off when it is not in use and take out 

the battery in case it will not be used for a long time.

1. Maximum voltage between terminal input and 

COM: 1000V (except for 200mV, 250V)

2. μA mA terminal input protection: (CE)250mA 265V 

auto recovery fuse

3. 10A terminal input protection: (CE)F1 (10A H 

1000V) Fast type melted fuse Φ10.3x38mm

4. Resistance input protection: PTC/250V

5. Capacitance input protection: (CE) F2, F3

(0.5A H 250V) quick-blow fuse 

    

5x20mm

6. Frequency input protection: PTC/1000V

7. Temperature input protection: (CE)0.5A 1000V 

fuse

8.         terminal input protection: PTC/1000V

9.  hFE input protection: (CE)250mA 265V auto 

recovery fuse, F3 (0.5A H 1000V) Fast type melted 

fuse Φ6.35x31.8mm

10. Display: Full functional LCD, maximum reading is 

19999, refreshing 2-3 times per second

11. Range: Manual

12. Polarity display: Auto

13. Over-range indication: 1

14. Low battery indication:“        ”  

15. Operating temperature: 0~40

 (32

~104

 )

16. Storage temperature: -10~50

 (14

~122

)

17. Relative humidity: 0

~30

 ≤75%

30

~ 40

 

≤50%

18. Electromagnetic compatibility: Under a radio 

frequency field of 1V/m, overall accuracy = specified 

ac 5% of the range; No specified accuracy if 

the unit is put under a radio frequency field of more 

than 1V/m. 

19. Power supply: AC (External power adapter 

AC110V/ DC9V-200mA) or DC (Battery Size 

R14/1.5V×6 PCS)

20. Product size: (300×245×105) mm

21 .Product Net Weight: about 1500g (accessories 

not included)

22. Safety Compliance: IEC 61010: CATII 1000V 

1. 

Manual Range 

        Manual range indicator

2. 

Warning!  

                Warning indicator

3.                                 Low Battery

4.                                 Indicator for high voltage

5.                                 Indicator for negative reading

6.                                 Indicator of AC voltage or 

                                    current (No display for DC  

                                    voltage or current)

7.                                 Data hold activated

8.                                 Test of diode

9.                                 Continuity buzzer on

10. 

Decimal Number 

  Showing the readout

11. Measurement units

72-14630

72-14630

Overview

Unpacking Inspection

General Specifications

Functional Buttons

Operational Guide (See Diagram 1, 2, 3)

LCD Display

Functions

Safety Notes

The Tenma 72-14630   Bench  Top    Digital     Multimeter 

features a 4 1/2 digit  display  with  a  maximum reading  

of 19999. The  large  backlit  LCD  makes  measurements  

easy  to read. The  well-featured  meter  also  provides  

full  range overload  protection  and  dual  power sources 

(AC or DC battery).  The  unit  can  be  used  to  test 

AC/DC voltage, AC/DC   current,   resistance,   frequency,     

capacitance, temperature    (Celsius),   hFE  transistors,   

diode,    and continuity test.

This operating manual includes relevant precautions and 

safety information. Please read it carefully and strictly 

observe  all  the  Warnings  and  Safety  Notes  contained  in 

this manual.

Open the package and take out the unit. Check the 

following items carefully to see if there is any missing or 

damaged.  If  you  find  anything  missing  or  damaged, 

please contact your local dealer immediately.

●  Operating Manual                                     1 Copy

●  Test Lead                                                  1 pair

●  Short Test Lead with Alligator Clip            1 pair

●  K Type Temperature Probe                       1 piece 

    (Only applicable to test below 230

)

●  Socket Adaptor                                          1 piece

●  Power Adapter                                           1 piece

    (110VAC 60Hz (US) , 9VDC 200mA)

This unit is designed and produced in strict compliance 

with  IEC61010-1  code. As  a  unit  with  double  insulation 

over-voltage  protection,  it  complies  with  the  safety 

standard  contained  in  CAT  II  1000V  and  Pollution 

Degree  II  codes.  Any  failure  to  follow  the  operating 

instruction contained in this manual may impair or void 

the protection it has. 

1 . Before use, the unit and test leads should be closely 

inspected. If there appears to be any damage to either the 

meter or test leads, do not attempt to use either. Do not use 

the meter if the cover is removed, as it exposes the user to 

2.  If test leads are damaged, they must be replaced with 

the one having identical model or electrical specifications.

3.  Do not touch fingers  or  bare skin to any  bare  cables, 

connectors,   unused   terminals,  or  circuits  while  being 

tested.

4.  When the meter is used at an effective voltage over 60V 

in DC or 30V in AC, special care should be taken as there 

is a danger of electric shock.

5.  If the scope of the input value remains unknown, just 

switch to the maximum range for testing.

6.  No voltage or current exceeding the rated voltage or 

current indicated by the unit should be applied between 

test leads or between test lead and ground.

7.  The rotary switch should be placed in the right position 

and no changeover of range should be made during 

measurement so as to prevent the damage of the unit.

8. Do not use or store the unit in an  enviroment  with high    

temperatures,     high     humidity,    flammable   materials,  

or  excessive  exposure to  magnetic  or radioactive  field.

Unit of voltage: millivolt, volt

Unit of current: microampere, 

milliampere, ampere 

Unit of electrical resistance: Ohm, 

kilohms, megohm

Unit of electrical capacity: nanofarad, 

microfarad

Unit of frequency: Kilohertz

Unit of temperature: Celsius degree
Unit of triode amplification: times

Terminal Input

Explanation

DC voltage measurement

AC voltage measurement

Resistance measurement  

Frequency measurement  

Diode/continuity test using 

buzzer

(Use socket adaptor)

(Use socket adaptor)

(Use socket adaptor)

mA/μA for DC current measurement
A for DC Current measurement
mA/μA for AC current measurement
A for AC current measurement  

Capacitance measurement

Temperature measurement

Power switch 

Press Hold to enter or exit the hold mode when in 

any mode, the meter beeps.

Backlight on or off (battery powered, backlight remain 

on about 10 seconds before auto shut down.) 

Triode amplification times 

measurement

the risk of electric shock.

damage