Appendix 19 Calibration
A41
A
ppend
ix
•
For more information about 0
Ω
calibration connections, see "Appendix 6
•
Adequate noise countermeasures must be implemented during high-resis-
tance and low-resistance measurement, when using the low measurement
current setting, and during low-power resistance measurement.
In a highly noisy environment, the measured value may become unstable or
inaccurate. In addition, the measurement error detection function may react
and no measured value may be displayed. Connect the metal exterior of
standard resistors and dial resistors to the instrument’s GUARD potential.
See:
"Appendix 7 Unstable Measured Values" (p.A12)
•
Do not use alligator clips with the voltage detection terminals. Thermal
EMFs may cause measured values to diverge.
TEMP.SENSOR
FLUKE
5520A
Hi
Lo
HIOKI
RM3545
Wire resistance: 500 m
Ω
or less
(circuit resistance)
(No polarity)
Temperature measurement (Thermistor)
TEMP.
ANALOG
INPUT
HIOKI
SS7012
+
−
HIOKI
RM3545
Wire resistance: 100
Ω
or less
(circuit resistance)
Temperature (Analog input)
+
−
D/A OUTPUT
HIOKI
3237
V
HIOKI
RM3545
Wire resistance: 100
Ω
or less
(circuit resistance)
D/A output
+
GND
COM
Summary of Contents for RM3545
Page 2: ......
Page 4: ......
Page 26: ...Operating Precautions 18...
Page 42: ...1 5 Checking the Measurement Target 34...
Page 54: ...2 6 Pre Operation Inspection 46...
Page 70: ...3 5 Checking Measured Values 62...
Page 226: ...10 6 Supplied Connector Assembly 218...
Page 290: ...Chapter 13 Specifications 282...
Page 312: ...14 4 Disposing of the Instrument 304...
Page 349: ...Appendix 18 Outline Drawing A37 Appendix Appendix 18 Outline Drawing...
Page 362: ...Index Index 4...
Page 363: ......
Page 364: ......