14.1 Troubleshooting
284
The following tables provide information about general issues. For more information about
issues related to measured values, the multiplexer, or the instrument’s external interfaces,
see the following pages.
If you are unable to find information about a particular issue, please contact your distributor.
1. General issues
14.1 Troubleshooting
Q&A (Frequently Asked Questions)
No
Issue
Items to check
Possible causes
→
Solutions
See
1-1 The instrument
cannot be turned
on. (The display
shows nothing.)
Color of the
STANDBY key
Green
The display settings have not been
configured correctly.
→
Adjust the backlight brightness and
contrast.
Red
The instrument is in the standby
state.
→
Press the STANDBY key.
None
(Off)
The instrument is not receiving
power.
→
Check the continuity of the power
cord.
→
Verify that a circuit breaker has not
been tripped.
→
Turn on the main power switch (on
the back of the instrument).
The supply voltage or frequency is
incorrect.
→
Check the power supply ratings
(100 to 240 V, 50/60 Hz).
1-2 The keys are
unresponsive.
Display
LOCK is shown. The key lock function is active.
→
Cancel the key lock function.
→
Turn OFF the EXT I/O KEY_LOCK
signal.
RMT is shown.
The instrument is in the remote state.
→
Cancel the remote state.
Neither LOCK
nor RMT is
shown.
Certain functions cannot be used
simultaneously.
→
See the list of functional limitations.
1-3 The instrument’s
comparator lamp
will not turn on.
Measured values Displayed
The comparator function is OFF.
→
Turn ON the function.
Not displayed
(Display other
than value)
If the measured value is not being
displayed, no judgment will be made,
and the lamp will not turn on.
−
Summary of Contents for RM3545
Page 2: ......
Page 4: ......
Page 26: ...Operating Precautions 18...
Page 42: ...1 5 Checking the Measurement Target 34...
Page 54: ...2 6 Pre Operation Inspection 46...
Page 70: ...3 5 Checking Measured Values 62...
Page 226: ...10 6 Supplied Connector Assembly 218...
Page 290: ...Chapter 13 Specifications 282...
Page 312: ...14 4 Disposing of the Instrument 304...
Page 349: ...Appendix 18 Outline Drawing A37 Appendix Appendix 18 Outline Drawing...
Page 362: ...Index Index 4...
Page 363: ......
Page 364: ......