Chapter 13 Specifications
265
13
(18)Comparator Function
* When set using the instrument’s keys, the input range will reflect the range and scaling coefficient with a max-
imum resolution of 1 n
Ω
and a maximum value of 9 G
Ω
.
Operation
Compares setting and measured values
Setting
ON/OFF (fixed range when the comparator function is on; the comparator func-
tion is automatically turned off when the
Δ
T and BIN functions are on)
Comparator mode
ABS mode/ REF% mode
Default state
OFF, ABS mode
Judgment
Hi
Measured value > Upper limit value
IN
Upper limit value
≥
measured value
≥
Lower limit value
Lo
Lower limit value > measured value
Total judgment function (RM3545-02 only)
Operation
When using the MUX measurement terminal setting with the scan function set
to AUTO or STEP, a PASS/FAIL judgment is made for each channel, and a to-
tal judgment is determined.
PASS/FAIL judgment
(for each scan channel)
PASS
When the comparator judgment satisfies the PASS conditions
FAIL
When the comparator judgment does not satisfy the PASS condi-
tions
Total judgment
PASS
When all channels are PASS or when the PASS condition is OFF
FAIL
When any channel is FAIL
PASS conditions
OFF/ Hi/ IN/ Lo/ Hi or Lo/ ALL (for each scan channel)
Default setting
IN
ABS Mode
Upper/Lower limit
ranges
0.000 0 m
Ω
to 9000.00 M
Ω
*
Default setting
0.000 0 m
Ω
REF% Mode
Display
Absolute value display and Relative value display
Relative value display
range
-999.999% to 999.999%
Reference value range
0.000 1 m
Ω
to 9000.00 M
Ω
*
When using the MUX measurement terminal setting, the measurement results
for scan channel 1 can be used as the reference value. (RM3545-02 only)
Upper/ Lower limit
ranges
0.000% to ±99.999%
Default setting
Reference value: 0.000 1 m
Ω
, Upper/ Lower limit ranges: 0.000%
(
)
(
)
(
)
[ ]
%
1 0 0
1
基 準 値
測 定 値
相 対 値
×
−
=
(Relative value) =
(Measured value)
(Reference value)
× 100 [%]
−
1
Summary of Contents for RM3545
Page 2: ......
Page 4: ......
Page 26: ...Operating Precautions 18...
Page 42: ...1 5 Checking the Measurement Target 34...
Page 54: ...2 6 Pre Operation Inspection 46...
Page 70: ...3 5 Checking Measured Values 62...
Page 226: ...10 6 Supplied Connector Assembly 218...
Page 290: ...Chapter 13 Specifications 282...
Page 312: ...14 4 Disposing of the Instrument 304...
Page 349: ...Appendix 18 Outline Drawing A37 Appendix Appendix 18 Outline Drawing...
Page 362: ...Index Index 4...
Page 363: ......
Page 364: ......