10.2 Timing Chart
198
(3) Scan function: Step
After the trigger, processing switches to the next channel and measurement is performed.
The total judgment (T_PASS, T_FAIL, T_ERR) signals are only output once measurement
of the last channel is complete.
•
Once the TRIG signal turns on after measurement of all channels is complete, measure-
ment will start again with the first channel.
•
During scanning, the TRIG, CAL, and 0ADJ signals are ignored without being held.
•
For channels for which an externally connected device is selected, EOM will turn on after
switching processing completes.
SCN_STOP operation
State
Judgment result/
BCD
Judgment result/ BCD: HI, IN, LO, ERR, PASS, FAIL, BCDm-n, RNG_OUT0 to 3
In this example, channels 1 through 20 have been set to ON.
Switching time
Switching
processing
Channel 1
Measure-
ment
Switching
processing
Channel 20
Measure-
ment
Switching
processing
Channel 2
Measure-
ment
Switching
processing
Channel 1
Measure-
ment
Switching
processing
Switching
processing
Channel 2
Measure-
ment
Reset
processing
Switching
processing
Channel 1
Measure-
ment
Channel 1
Measure-
ment
State
Judgment result/
BCD
Approx.
15 ms
Approx.
10 ms
Summary of Contents for RM3545
Page 2: ......
Page 4: ......
Page 26: ...Operating Precautions 18...
Page 42: ...1 5 Checking the Measurement Target 34...
Page 54: ...2 6 Pre Operation Inspection 46...
Page 70: ...3 5 Checking Measured Values 62...
Page 226: ...10 6 Supplied Connector Assembly 218...
Page 290: ...Chapter 13 Specifications 282...
Page 312: ...14 4 Disposing of the Instrument 304...
Page 349: ...Appendix 18 Outline Drawing A37 Appendix Appendix 18 Outline Drawing...
Page 362: ...Index Index 4...
Page 363: ......
Page 364: ......