4.4 Stabilizing Measured Values (Averaging Function)
73
4
The averaging function averages multiple measured values and displays the results. It can
be used to reduce variation in measured values.
For internal trigger measurement (Free-Run), a moving average is calculated.
For external trigger measurement (and
:READ?
command operation) (Non-Free-Run), a
mean average is used.
For more information about communications commands, see the included application disc.
Average (of measurements D1 to D6) with Averaging Samples set to 2.
When using the SLOW2 measurement speed with low-power resistance measurement on,
the instrument will performing averaging with two iterations internally even if the averaging
function is set to off. If the averaging function is on, the instrument will perform averaging
using the set number of iterations.
4.4
Stabilizing Measured Values
(Averaging Function)
1st Sample
2nd Sample
3rd Sample
Free-Run (Moving Avg.)
(D1+D2)/2
(D2+D3)/2
(D3+D4)/2
Non-Free-Run (Mean Avg.)
(D1+D2)/2
(D3+D4)/2
(D5+D6)/2
1
Open the Settings Screen.
2
Open the Measurement Setting Screen.
The Settings screen
appears.
Switch the function menu
to P.2/3.
1
2
Move the cursor to the
[MEAS] tab with the left and
right cursor keys.
Summary of Contents for RM3545
Page 2: ......
Page 4: ......
Page 26: ...Operating Precautions 18...
Page 42: ...1 5 Checking the Measurement Target 34...
Page 54: ...2 6 Pre Operation Inspection 46...
Page 70: ...3 5 Checking Measured Values 62...
Page 226: ...10 6 Supplied Connector Assembly 218...
Page 290: ...Chapter 13 Specifications 282...
Page 312: ...14 4 Disposing of the Instrument 304...
Page 349: ...Appendix 18 Outline Drawing A37 Appendix Appendix 18 Outline Drawing...
Page 362: ...Index Index 4...
Page 363: ......
Page 364: ......