10.1 External Input/Output Connector and Signals
181
10
(3) Output Signals
•
When not displaying the Measurement screen and while error messages (except Setting
Monitor errors) are being displayed, input signals are disabled. However, the PRINT sig-
nal is valid on the Statistical Calculation Results screen.
•
EXT I/O input and output signals are not usable while changing measurement settings.
EOM
This signal indicates the end of measurement and zero-adjustment. At this
point in time, the comparator judgment results and the ERR, BCD and BIN
signals have been finalized.
INDEX
This signal indicates that A/D conversion in the measurement circuit is fin-
ished.
When the asserted (ON) state occurs, the measurement target can be
removed.
ERR
This signal indicates that a measurement fault has occurred (except out-of-
range detection). It is updated simultaneously with the EOM signal. At this
time, comparator judgment outputs are all de-asserted (OFF).
HI, IN, LO
These are the comparator judgment output signals.
HILO
When using BCD output, pin 11 outputs the result of an OR operation applied
to the Hi and Lo judgments.
T_PASS,
T_FAIL,
T_ERR
These are the total judgment results. They are valid only when the scan func-
tion is set to auto or step.
BCDm-n
When using BCD output, this signal outputs n bits of digit m. (When BCD1-x is
the lowermost digit, BCDX-0 is the LSB.)
When the measured value display is “
OvrRng
”, “
CONTACT TARM
”, or “
- - - -
- - -
,” all digits of BCD output will be 9.
When the measured value display is a negative value, all digits of BCD output
will be 0. When the lower limit value has been set to 0 and a negative mea-
sured value is encountered, the LO signal will be output in accordance with
the display screen result. However, when using the comparator's REF%
mode, an unsigned value equivalent to the absolute value being displayed
(i.e., an absolute value) will be output.
OB,
BIN0 to BIN9
When BIN output has been configured, the BIN judgment results will be output
from pins 13 to 17 and pins 31 to 36. When the results do not correspond to
BIN0 to BIN9, OB will turn on.
OUT0 to
OUT2
When the output mode is judgment mode, pins 18, 19 and 37 can be used as
general-purpose output pins. The output signals can be controlled with the
:IO:OUT
command.
See:
Communications Command Instruction Manual on the included applica-
tion disc.
RNG_OUT0
to
RNG_OUT3
When BCD_LOW is turned ON when using BCD output, range information
can be acquired from pins 12, 13, 29, and 31.
Summary of Contents for RM3545
Page 2: ......
Page 4: ......
Page 26: ...Operating Precautions 18...
Page 42: ...1 5 Checking the Measurement Target 34...
Page 54: ...2 6 Pre Operation Inspection 46...
Page 70: ...3 5 Checking Measured Values 62...
Page 226: ...10 6 Supplied Connector Assembly 218...
Page 290: ...Chapter 13 Specifications 282...
Page 312: ...14 4 Disposing of the Instrument 304...
Page 349: ...Appendix 18 Outline Drawing A37 Appendix Appendix 18 Outline Drawing...
Page 362: ...Index Index 4...
Page 363: ......
Page 364: ......