Chapter 13 Specifications
267
13
(22)Temperature Conversion Function (
Δ
T)
(23)Statistical Calculations
Operation
Utilizing the temperature-dependent nature of resistance, the temperature con-
version function converts resistance measurements for display as tempera-
tures.
Formula
Δ
T display range
-9999.9 to 9999.9°C
Setting
ON/ OFF
(Fixed when the comparator function is off;
Δ
T is automatically turned off when
TC, the statistical calculation function, and the BIN function are on.)
Default setting
OFF,
t
1
: 23.0°C,
R
1
: 1.000 0
Ω
,
k
: 235.0
Δ
t
: Temperature increase (°C)
t
1
: Winding temp. (°C, cool state) when measuring initial resistance
R
1
Setting range: -10.0 to 99.9°C
t
2
: Ambient temp. (°C) at final measurement
R
1
: Winding resistance (
Ω
) at temp.
t
1
(cool state)
Setting range: 0.001
μΩ
to 9000.000 M
Ω
*
R
2
: Winding resistance (
Ω
) at final measurement
k
: Reciprocal (°C) of temp. coefficient of conductor material at 0°C
Setting range: -999.9 to 999.9
* When set using the instrument’s keys, the input range will reflect the range and scaling coefficient with
a maximum resolution of 1 n
Ω
and a maximum value of 9 G
Ω
.
)
(
)
(
2
1
1
2
t
k
t
k
R
R
t
+
−
+
=
Δ
Operation
Statistical calculations are performed on measured values.
Setting
ON/ OFF (The statistical calculation function is automatically turned off when
Δ
T is on while using the MUX measurement terminal setting.)
Maximum number of
data points
30,000
Calculations
Total data count, Number of valid data samples, Mean, Minimum value (index
no.), Maximum value (index no.), Standard deviation of sample,
Population standard deviation
• When the comparator function is ON
Count for each comparator judgment, Process capability indices (disper-
sion, bias)
• When the BIN function is ON
Count for each BIN number, OUT (Hi or Lo) count for all BIN numbers,
invalid BIN count
Clearing calculations
Clear all data / clear 1 data point (to revert to data immediately before mea-
surement)
Default setting
OFF
Summary of Contents for RM3545
Page 2: ......
Page 4: ......
Page 26: ...Operating Precautions 18...
Page 42: ...1 5 Checking the Measurement Target 34...
Page 54: ...2 6 Pre Operation Inspection 46...
Page 70: ...3 5 Checking Measured Values 62...
Page 226: ...10 6 Supplied Connector Assembly 218...
Page 290: ...Chapter 13 Specifications 282...
Page 312: ...14 4 Disposing of the Instrument 304...
Page 349: ...Appendix 18 Outline Drawing A37 Appendix Appendix 18 Outline Drawing...
Page 362: ...Index Index 4...
Page 363: ......
Page 364: ......