4.12 Maintaining Measurement Precision (Self-Calibration)
95
4
Manual setting operation
Self-calibration starts immediately when the CAL signal is input.
If the TRIG signal is input during self-calibration, self-calibration will continue. In this case,
the TRIG signal will be accepted, the EOM signal will turn off, and measurement will start
after self-calibration completes.
If the CAL signal is received during measurement, the CAL signal will be accepted, and
self-calibration will start after measurement completes.
If the TRIG signal is received during self-calibration
Method of normal use
Self-Calibration
Measurement
processing
Measurement
Measurement
Measurement
processing
Measurement
Measurement
Self-Calibration