8.4 Measuring with the Multiplexer
162
This section describes how to measure channels in successive order.
Before configuring these settings, see "Configuring Multiplexer Settings" (p.148) and "Cus-
tomizing Measurement Conditions for Individual Channels" (p.160).
During scanning measurement, only the Standby and
[STOP]
keys can be used.
Performing Scan Measurement
1
Set the scan function to either auto or step.
See:
"Configuring Multiplexer Settings" (p.148)
2
Input the external trigger to perform measurement. (trigger input: EXT I/O
TRIG signal, ENTER (trigger) key,
∗
TRG
command)
The measurement results will be displayed.
Scanning can be stopped by pressing the
[STOP]
key during scanning.
•
When the scan function is set to auto
Scanning will stop midway through the scan.
•
When the scan function is set to step
If there is a scan in progress, it will return to the first channel.
When the scan function is set to step, you will need to input the trigger for each
channel. When the scan function is set to auto, you can measure all channels
with a single trigger input.
•
When the scan function is set to auto or step, the trigger source will be set to
an external trigger ([EXT]).
•
When the scan function is auto or step, the range, comparator, and speed
cannot be changed on the Measurement screen. Instead, these settings
must be changed on the Settings screen.
•
When the scan function is set to AUTO, channels that are set to an externally
connected device will be ignored.
Channel number
Total judgment result
Individual channels’ measured value
Individual channels’ comparator result
(If a measured value fault occurs, a description of the error is displayed.)
Individual channels’
PASS/FAIL judgment re-
sult (If a measured value
fault occurs, ”
---
” is dis-
played.)
Summary of Contents for RM3545
Page 2: ......
Page 4: ......
Page 26: ...Operating Precautions 18...
Page 42: ...1 5 Checking the Measurement Target 34...
Page 54: ...2 6 Pre Operation Inspection 46...
Page 70: ...3 5 Checking Measured Values 62...
Page 226: ...10 6 Supplied Connector Assembly 218...
Page 290: ...Chapter 13 Specifications 282...
Page 312: ...14 4 Disposing of the Instrument 304...
Page 349: ...Appendix 18 Outline Drawing A37 Appendix Appendix 18 Outline Drawing...
Page 362: ...Index Index 4...
Page 363: ......
Page 364: ......