5.4 Performing Temperature Rise Test (Temperature Conversion Function (
Δ
T))
118
Guideline for k
IEC 60034 recommends the following:
•
Copper: k = 235
•
Aluminum: k = 225
See:
"Appendix 5 Temperature Conversion (
5
Set the reciprocal (k) of the temperature coefficient at 0°C.
6
Return to the Measurement screen.
Setting range: -999.9 to 999.9 (default: 235.0)
Move the cursor to the setting you
wish to configure. Make the value
editable with the
key.
2
Move the cursor to the digit you
wish to set with the left and right
cursor keys. Change the value
with the up and down cursor keys.
Change
values.
1
Move among
digits.
3
Accept
( Cancel)
Return to the
Measurement screen.