10.2 Timing Chart
201
10
Measured value (BCD) acquisition processing when using an external
trigger
For BCD output, the upper and lower digits must be acquired separately. The upper and
lower digits can be acquired in any order. In the following example, the upper digits are
acquired first.
Instrument
Controller
Measurement
start
Judgment re-
sults OFF
Measurement
start
Measurement
Measurement
complete
Reception of
measurement
complete
Acquisition of
BCD upper digits
Specification of
BCD lower digits
Acquisition of
BCD lower digits
BCD lower digit
setup start
BCD lower digit
setup complete
BCD4-0 to
BCD7-3
BCD1-0 to
BCD3-3
Wait for EOM
(when detect-
ing the level,
after a 0.5 ms
wait)
Setup time: Max. 0.6 ms
(BCD upper setup time: Max. 0.4 ms)