4.5 Correcting for the Effects of Temperature (Temperature Correction (TC))
76
4
Set the reference temperature and temperature coefficient.
(Set the reference temperature and temperature coefficient by following steps
1
through
3
for each.)
5
Return to the Measurement screen.
Setting range
reference temperature : -10.0 to 99.9°C (default: 20°C)
temperature coefficient : -99999 to 99999ppm/°C (default: 3930ppm/°C)
Move the cursor to the setting you
wish to configure. Make the value
editable with the
key.
2
Move the cursor to the digit you
wish to set with the left and right
cursor keys. Change the value
with the up and down cursor keys.
Change
values.
1
Move among
digits.
3
Accept
( Cancel)
Return to the
Measurement screen.