4.6 Correcting Measured Values and Displaying Physical Properties Other than Resistance
80
Scaling calculation is performed on measured values after zero-adjustment calculation.
Consequently, measured values may not equal zero even after zero adjustment.
•
If the calculation result exceeds the display range, the measured value will not be dis-
played at full scale.
Example: If you set an offset of 90
Ω
for the 10
Ω
range
→
Values in excess of 10
Ω
will be displayed as OvrRng.
•
If the calculation result is negative, the displayed value will be negative.
Example: If you set an offset of -50 m
Ω
for the 100 m
Ω
range
→
If you measure 30 m
Ω
, -20 m
Ω
will be displayed.
6
Set the units for the displayed measured values.
7
Edit the unit as desired.
8
Return to the Measurement screen.
2
Selection
1
Use
Ω
as the unit. (default)
(go to step 8)
Eliminate the unit.
(go to step 8)
Use a user-defined unit.
Enter a number from 0 to 9
Enter a letter from A to z
Delete 1 character.
1
Move the cursor to the digit you
wish to set with the left and right
cursor keys. Change the value
with the up and down cursor keys.
Change
values.
Move among
digits.
2
Accept
( Cancel)
Make the value editable with the
key.
Return to the
Measurement screen.
Summary of Contents for RM3545
Page 2: ......
Page 4: ......
Page 26: ...Operating Precautions 18...
Page 42: ...1 5 Checking the Measurement Target 34...
Page 54: ...2 6 Pre Operation Inspection 46...
Page 70: ...3 5 Checking Measured Values 62...
Page 226: ...10 6 Supplied Connector Assembly 218...
Page 290: ...Chapter 13 Specifications 282...
Page 312: ...14 4 Disposing of the Instrument 304...
Page 349: ...Appendix 18 Outline Drawing A37 Appendix Appendix 18 Outline Drawing...
Page 362: ...Index Index 4...
Page 363: ......
Page 364: ......