5.2 Classifying Measurement Results (BIN Measurement Function)
108
BIN Measurement compares a measured value with up to ten sets of upper and lower
thresholds (BIN 0 to BIN 9) in one operation, and display the results. Measured values that
do not fall in any BIN are judged to be OB (out-of-bin). Judgment results are output at the
EXT I/O connector.
See:
"Connector Type and Signal Pinouts" (p. 177)
•
When the BIN measurement function is on, the comparator cannot be turned on.
•
Turning on
Δ
T or setting the measurement terminal to multiplexer automatically turns off
the BIN measurement function.
•
The range cannot be changed while using the BIN measurement function. To change the
range, do so with the
and
keys on the BIN Number Settings screen.
Turn off the BIN measurement function when using auto-ranging.
5.2
Classifying Measurement Results
(BIN Measurement Function)
1
Open the Settings Screen.
2
Open the BIN Setting Screen.
3
Open the BIN Setting Screen.
The Settings screen
appears.
Switch the function menu
to P.2/3.
1
2
Move the cursor to the [BIN]
tab with the left and right cur-
sor keys.
Enables the BIN function
Disables the BIN function
(default)
Selection
2
1
Summary of Contents for RM3545
Page 2: ......
Page 4: ......
Page 26: ...Operating Precautions 18...
Page 42: ...1 5 Checking the Measurement Target 34...
Page 54: ...2 6 Pre Operation Inspection 46...
Page 70: ...3 5 Checking Measured Values 62...
Page 226: ...10 6 Supplied Connector Assembly 218...
Page 290: ...Chapter 13 Specifications 282...
Page 312: ...14 4 Disposing of the Instrument 304...
Page 349: ...Appendix 18 Outline Drawing A37 Appendix Appendix 18 Outline Drawing...
Page 362: ...Index Index 4...
Page 363: ......
Page 364: ......