5.1 Judging Measured Values (Comparator Function)
103
5
When REF% mode is enabled, the measured value will be displayed as an absolute value
(%).
Example setting: Set a reference value of 10 m
Ω
with
±
1% decision tolerance.
To abort the setting process, press
. Settings are abandoned and the display returns to the pre-
vious screen.
Decide According to Reference Value and Tolerance (REF% Mode)
1
Open the relative tolerance setting screen.
2
Set the range.
3
Set the reference value.
Pressing an inoperative key during setting sounds a low-pitch beep (when the key beeper is enabled).
Relative Value =
(tolerance)
Reference
Value
Measured
Value
- 1
× 100 [%]
Setting range: -99.999% to +99.999%
Reference value
Allowable range (up-
per and lower limit
ranges relative to the
reference value)
Select the range you wish to use.
Change the decimal point position and unit (changes each time you
press the button).
Move the cursor to the digit you
wish to set with the left and right
cursor keys. Change the value
with the up and down cursor keys.
Change
values.
Move among
digits.
To Reset Numerical Values
Press
to clear the reference
value. The reference value will be re-
set to 0.
When using REF% mode and the multiplexer, the CH1 measurement results can be used
as the reference value by pressing
on MENU P.2/2.
Summary of Contents for RM3545
Page 2: ......
Page 4: ......
Page 26: ...Operating Precautions 18...
Page 42: ...1 5 Checking the Measurement Target 34...
Page 54: ...2 6 Pre Operation Inspection 46...
Page 70: ...3 5 Checking Measured Values 62...
Page 226: ...10 6 Supplied Connector Assembly 218...
Page 290: ...Chapter 13 Specifications 282...
Page 312: ...14 4 Disposing of the Instrument 304...
Page 349: ...Appendix 18 Outline Drawing A37 Appendix Appendix 18 Outline Drawing...
Page 362: ...Index Index 4...
Page 363: ......
Page 364: ......