10.4 External I/O Settings
211
10
Filter function (TRIG signal example)
•
Using the ON edge
•
Using the OFF edge
Hold the input signal until the response time elapses.
4
Set the response time.
5
Return to the Measurement screen.
Setting range: 50 ms to 500 ms (default: 50 ms)
Move the cursor to the setting you
wish to configure. Make the value
editable with the
key.
2
Move the cursor to the digit you
wish to set with the left and right
cursor keys. Change the value
with the up and down cursor keys.
Change
values.
1
Move among
digits.
3
Accept
( Cancel)
Return to the
Measurement screen.
Measurement
processing
Measurement
TRIG
ON
OFF
Response time
(Chatter)
Measurement
TRIG
ON
OFF
(Chatter)
Measurement
processing
Response time