Appendix 19 Calibration
A39
A
ppend
ix
D/A output
Calibration points
*1 Without 0ADJ: Only with OVC off
Equipment
Calibration point
Manufacturer
Standard model
Voltmeter
0
Ω
: 0 V
HIOKI
Equivalent to 3237
1
Ω
: 1 V
Output cable
Wiring resistance: 500 m
Ω
or less
(circuit resistance)
Range
Calibration point
OVC
Measure
ment
Current
100 M
Ω
range high-
precision
mode
0ADJ
Resistance
measurement
(Low Power: OFF)
10 m
Ω
0
Ω
, 10 m
Ω
ON, OFF
-
-
With or
without
*1
100 m
Ω
0
Ω
, 100 m
Ω
ON, OFF
High, Low
-
With or
without
*1
1
Ω
0
Ω
, 1
Ω
ON, OFF
High, Low
-
With or
without
*1
10
Ω
0
Ω
, 10
Ω
ON, OFF
High, Low
-
With or
without
*1
100
Ω
0
Ω
, 100
Ω
ON, OFF
High, Low
-
With or
without
*1
1000
Ω
0
Ω
, 1 k
Ω
ON, OFF
-
-
With or
without
*1
10 k
Ω
0
Ω
, 10 k
Ω
OFF
-
-
-
100 k
Ω
0
Ω
, 100 k
Ω
OFF
-
-
-
1000 k
Ω
0
Ω
, 1 M
Ω
OFF
-
-
-
10 M
Ω
0
Ω
, 10 M
Ω
OFF
-
-
-
100 M
Ω
0
Ω
, 100 M
Ω
OFF
-
ON, OFF
-
1000 M
Ω
0
Ω
, 1000 M
Ω
OFF
-
OFF
-
Resistance
measurement
(Low Power: ON)
1000 m
Ω
0
Ω
, 1
Ω
ON
-
-
-
10
Ω
0
Ω
, 10
Ω
ON
-
-
-
100
Ω
0
Ω
, 100
Ω
ON
-
-
-
1000
Ω
0
Ω
, 1 k
Ω
ON
-
-
-
Temperature
(thermistor)
25°C: 2186.0
Ω
input
Temperature
(analog input)
10°C: 0.1 V input
100°C: 1 V input
D/A output
1
Ω
0
Ω
: 0 V input
1
Ω
: 1 V input
Summary of Contents for RM3545
Page 2: ......
Page 4: ......
Page 26: ...Operating Precautions 18...
Page 42: ...1 5 Checking the Measurement Target 34...
Page 54: ...2 6 Pre Operation Inspection 46...
Page 70: ...3 5 Checking Measured Values 62...
Page 226: ...10 6 Supplied Connector Assembly 218...
Page 290: ...Chapter 13 Specifications 282...
Page 312: ...14 4 Disposing of the Instrument 304...
Page 349: ...Appendix 18 Outline Drawing A37 Appendix Appendix 18 Outline Drawing...
Page 362: ...Index Index 4...
Page 363: ......
Page 364: ......