8.3 Multiplexer Settings
149
8
3
Set the measurement terminals.
4
Select the measurement method.
When the measurement method is switched, the multiplexer channel settings will be
initialized (i.e., a multiplexer channel reset will be performed). Always finalize the
measurement method before allocating pins or performing zero-adjustment.
5
Set the scan function.
When the scan function is set to auto or step, external trigger operation will be used
regardless of the trigger source setting.
2
Make measurements using
the front measurement ter-
minals (do not use the multi-
plexer). (default)
Use the multiplexer.
Selection
1
2
4-wire (default)
2-wire
Selection
1
2
Use auto-scan
(measure all channels at
each TRIG signal).(default)
Use step scan
(measure 1 channel at each
TRIG signal).
Do not scan.
Selection
1