10.2 Timing Chart
194
For more information about the contact improver function, see p.90.
Probe contacts can be improved by applying current between the sense terminals before
measurement.
The maximum contact improvement current is 10 mA, and the maximum applied voltage is
5 V. When low power is set to on, the contact improver function is set to off. Using the con-
tact improver function causes the time until measurement completion to be lengthened by 2
ms.
Timing Chart (Contact Improver Function)
Contact improver timing
The Contact Improver function applies voltage to the sample. Be careful
when measuring samples with characteristics that may be affected.
Probe contact condition
Contact Improver current
Measurement current
(10 m
Ω
to 1000
Ω
range)
Measurement current
(10 k
Ω
to 1000 M
Ω
range)
Measurement
Delay
Measuring