
4.11 Improving Probe Contact (Contact Improver Function)
90
Probe contacts can be improved by applying current from the SENSE A to the SENSE B
probes before measuring.
The maximum contact improvement current is 10 mA, and the maximum applied voltage is 5 V.
When low power is set to on, the contact improver function is set to off.
Using the contact improver function causes the time until measurement completion to be
lengthened by 0.2 ms.
Timing Chart (Contact Improver Function)
4.11 Improving Probe Contact (Contact Improver
Function)
The Contact Improver function applies voltage to the sample. Be careful
when measuring samples with characteristics (magnetoresistive elements,
signal relays, EMI filters, etc.) that may be affected.
1
Open the Settings Screen.
Probe contact condition
Contact Improver current
Measurement current
(10 m
Ω
to 1,000
Ω
range)
Measurement current
(10 k
Ω
to 1,000 M
Ω
range)
Measurement
Delay
Measuring
The Settings screen
appears.
Switch the function menu
to P.2/3.
1
2
Summary of Contents for RM3545
Page 2: ......
Page 4: ......
Page 26: ...Operating Precautions 18...
Page 42: ...1 5 Checking the Measurement Target 34...
Page 54: ...2 6 Pre Operation Inspection 46...
Page 70: ...3 5 Checking Measured Values 62...
Page 226: ...10 6 Supplied Connector Assembly 218...
Page 290: ...Chapter 13 Specifications 282...
Page 312: ...14 4 Disposing of the Instrument 304...
Page 349: ...Appendix 18 Outline Drawing A37 Appendix Appendix 18 Outline Drawing...
Page 362: ...Index Index 4...
Page 363: ......
Page 364: ......