10.2 Timing Chart
189
10
BCDm_n signal transition time based on the BCD_LOW signal
Internal wait time (unit: ms) (Processing time before and after integration measure-
ment) reference values
LP OFF
Range
100 M
Ω
range
high-precision mode
Measurement
Current
Time
10 m
Ω
−
−
40
100 m
Ω
−
High
40
−
Low
1.8
1000 m
Ω
−
High
1.5
−
Low
1.3
10
Ω
−
High
1.5
−
Low
1.3
100
Ω
−
High
2.1
−
Low
1.3
1000
Ω
−
−
2.3
10 k
Ω
−
−
12
100 k
Ω
−
−
20
1000 k
Ω
−
−
150
10 M
Ω
−
−
570
100 M
Ω
ON
−
1300
OFF
−
300
1000 M
Ω
OFF
−
400
LP ON
Range
Time
1000 m
Ω
15
10
Ω
35
100
Ω
35
1000
Ω
36
BCD Signal Timing
BCD4-0 to BCD7-3
BCD4-0 to BCD7-3
BCD1-0 to BCD3-3