Chapter 13 Specifications
262
(14)Scaling Function
Operation
Measured values are corrected with the linear function
R
S
=
A
×
R
+
B
R
S
: Value after scaling
A
: Gain coefficient
Setting range: 0.200 0 × 10
-3
to 1.999 9 × 10
3
R
: Measured value after zero-adjustment and temperature correction
B
: Offset
Setting range: 0 to ±9 × 10
9
(maximum resolution: 1 n
Ω
)
Setting
ON/ OFF
Display format
See below. (When 9 G is exceeded, the over-range display is shown.)
Unit
Ω
/ none/ user-selected 3 characters (Except SI prefix)
Default setting
OFF,
A
: 1.0000 ×1,
B
: 0, Unit:
Ω
Low-Power: OFF
* When high-precision mode is off in the 100 M
Ω
range, 5 digits are displayed.
Low-Power: ON
Range
Gain coefficient
(0.2000 to
1.9999)
×10
-3
(0.2000 to
1.9999)
×10
-2
(0.2000 to
1.9999)
×10
-1
(0.2000 to
1.9999)
×1(10
0
)
(0.2000 to
1.9999)
×10(10
1
)
(0.2000 to
1.9999)
×10
2
(0.2000 to
1.9999)
×10
3
10 m
Ω
00.000
μ
000.000
μ
0000.000
μ
00.000 00 m
000.000 0 m
0000.000
m 00.000 00
100 m
Ω
000.000
μ
0000.000
μ
00.000 00 m 000.000 0 m
0000.000
m
00.000 00
000.000 0
1000 m
Ω
0000.000
μ
00.000 00 m 000.000 0 m 0000.000
m
00.000 00
000.000 0
0000.000
10
Ω
00.00 000 m 000.000 0 m 0000.000
m 00.000 00
000.000 0
0000.000
00.000 00 k
100
Ω
000.000 0 m 0000.000
m 00.000 00
000.000 0
0000.000
00.000 00 k
000.000 0 k
1000
Ω
0000.000
m 00.000 00
000.000 0
0000.000
00.000 00 k
000.000 0 k
0000.000
k
10 k
Ω
00.000 00
000.000 0
0000.000
00.000 00 k
000.000 0 k
0000.000
k
00.000 00 M
100 k
Ω
000.000 0
0000.000
00.000 00 k
000.000 0 k
0000.000
k
00.000 00 M 000.000 0 M
1000 k
Ω
0000.000
00.000 00 k
000.000 0 k
0000.000
k
00.000 00 M
000.000 0 M 0000.000
M
10 M
Ω
00.000 00 k
000.000 0 k
0000.000
k
00.000 00 M
000.000 0 M
0000.000
M 00.000 00 G
100 M
Ω
* 000.000 0 k
0000.000
k
00.000 00 M 000.000 0 M
0000.000
M
00.000 00 G 000.000 0 G
1000 M
Ω
0000.0
k
00.000
M 000.00
M 0000.0
M
00.000
G
000.00
G 0000.0
G
Range
Gain coefficient
(0.2000 to
1.9999)
×10
-3
(0.2000 to
1.9999)
×10
-2
(0.2000 to
1.9999)
×10
-1
(0.2000 to
1.9999)
×1(10
0
)
(0.2000 to
1.9999)
×10(10
1
)
(0.2000 to
1.9999)
×10
2
(0.2000 to
1.9999)
×10
3
1000 m
Ω
0000.00
μ
00.000 0
m 000.000
m 0000.00
m
00.000 0
000.000
0000.00
10
Ω
00.000 0
m 000.000
m 0000.00
m 00.000 0
000.000
0000.00
00.000 0
k
100
Ω
000.000
m 0000.00
m 00.000 0
000.000
0000.00
00.000 0
k
000.000
k
1000
Ω
0000.00
m 00.000 0
000.000
0000.00
00.000 0
k
000.000
k
0000.00
k
Summary of Contents for RM3545
Page 2: ......
Page 4: ......
Page 26: ...Operating Precautions 18...
Page 42: ...1 5 Checking the Measurement Target 34...
Page 54: ...2 6 Pre Operation Inspection 46...
Page 70: ...3 5 Checking Measured Values 62...
Page 226: ...10 6 Supplied Connector Assembly 218...
Page 290: ...Chapter 13 Specifications 282...
Page 312: ...14 4 Disposing of the Instrument 304...
Page 349: ...Appendix 18 Outline Drawing A37 Appendix Appendix 18 Outline Drawing...
Page 362: ...Index Index 4...
Page 363: ......
Page 364: ......