Chapter 13 Specifications
273
13
EOM output timing setting
Setting
HOLD/ PULSE
Operation
When using an external trigger source with the HOLD setting, the on state is
held until the next TRIG signal or 0ADJ signal is input.
When using an external trigger source with the PULSE setting, the off state is
held after the pulse width setting has elapsed.
When using the internal trigger source, EOM output is fixed to pulse output with
a width of 5 ms (when using auto self-calibration) or no EOM output is gener-
ated (when using manual self-calibration), regardless of the EOM output timing
setting.
Pulse width
1 ms to 100 ms
Default setting
HOLD, 5 ms
EXT I/O test function
Operation
Displays the EXT I/O input signal state and generates output signals as de-
sired.
External power output
Output voltage
Sink output: 5.0 V ±10%, source output: -5.0 V ±10%, 100 mA max.
Isolation
Floating from protective ground potential and measurement circuit
Insulation rating
Terminal to ground voltage:
Not more than 50 VDC, 33 Vrms AC, and 46.7 Vpk AC