10.2 Timing Chart
193
10
Manual setting operation
Self-calibration starts immediately when the CAL signal is input. If the TRIG signal is input
during self-calibration, self-calibration will continue. In this case, the TRIG signal will be
accepted, the EOM signal will turn off, and measurement will start after self-calibration com-
pletes. If the CAL signal is received during measurement, the CAL signal will be accepted,
and self-calibration will start after measurement completes.
If the TRIG signal is received during self-calibration
Method of normal use
Self-calibration
Measurement
processing
Measure-
ment
Measure-
ment
Measurement
processing
Measure-
ment
Measure-
ment
Self-calibration