5.1 Judging Measured Values (Comparator Function)
101
5
Setting example: Upper threshold 12 m
Ω
, lower threshold 10 m
Ω
To abort the setting process, press
. Settings are abandoned and the display returns to the pre-
vious screen.
Decide According to Upper/Lower Thresholds (ABS Mode)
1
Open the absolute value threshold setting screen.
2
Set the range.
3
Set the positive tolerance.
Upper threshold
Lower threshold
Switch to
ABS mode.
Select the range you wish to use.
Change the decimal point position and unit (changes each time you
press the button).
Move the cursor to the digit you
wish to set with the left and right
cursor keys. Change the value
with the up and down cursor keys.
Change
values.
Move among
digits.
To Reset Numerical Values
Press
to clear the upper limit
value. The upper limit value will be re-
set to 0.