Appendix 6 Zero Adjustment
A7
A
ppend
ix
Zero adjustment is a function which adjusts the zero point by deducting the residual value
obtained during 0
Ω
measurement. For this reason, zero adjustment must be performed
when connection is made to 0
Ω
. However, connecting a sample with no resistance is diffi-
cult and therefore is not practical.
In this respect, when performing the actual zero adjustment, create a pseudo connection to
0
Ω
and then adjust the zero point.
To create 0
Ω
connection state
If an ideal 0
Ω
connection is made, the voltage between SENSE A and SENSE B becomes
0 V according to the Ohm's Law of
E
=
I
×
R
. In other words, if you set the voltage between
SENSE A and SENSE B to 0 V, this gives you the same state of 0
Ω
connection.
To perform zero adjustment using the instrument
The instrument uses a measurement fault detection function to monitor the state of connec-
tion between measurement terminals. For this reason, when performing zero adjustment,
you need to make connections between the terminals appropriately in advance (Fig. 1).
First, short between SENSE A and SENSE B to set
the voltage between SENSE A and SENSE B to 0 V. If
lead resistances
R
SEA
and
R
SEB
of the cable are less
than few
Ω
, there will be no problem. Because the
SENSE terminal is a voltage measurement terminal,
almost no current
I
0
flows. Therefore, in the
E
=
I
0
×
(
R
SEA
+
R
SEB
) formula,
I
0
≈
0 is achieved; if lead resis-
tances
R
SEA
and
R
SEB
are less than few
Ω
, voltage
between SENSE A and SENSE B will become almost
zero.
Next, make connection between SOURCE A and
SOURCE B. This is to avoid display of error when no
measurement current flows through. Lead resistances
R
SOA
and
R
SOB
of the cable must be less than the resis-
tance for flowing measurement current.
Furthermore, if the instrument also monitors the con-
nection between SENSE and SOURCE, you need to make connection between SENSE
and SOURCE. If lead resistance
R
Short
of the cable has only few
Ω
, there will be no prob-
lem.
If you wire in the way described above, measurement current
I
flowing out from SOURCE B
will go to SOURCE A but not to the lead of SENSE A or SENSE B. This enables the voltage
between SENSE A and SENSE B to be kept accurately at 0 V, and appropriate zero adjust-
ment becomes possible.
Appendix 6
Zero Adjustment
E
= (
I
0
×
R
SEB
) + (
I
0
×
R
SEA
)
= (0 ×
R
SEB
) + (0 ×
R
SEA
)
= 0 [V]
Figure 1.
Pseudo connection to 0
Ω
Summary of Contents for RM3545
Page 2: ......
Page 4: ......
Page 26: ...Operating Precautions 18...
Page 42: ...1 5 Checking the Measurement Target 34...
Page 54: ...2 6 Pre Operation Inspection 46...
Page 70: ...3 5 Checking Measured Values 62...
Page 226: ...10 6 Supplied Connector Assembly 218...
Page 290: ...Chapter 13 Specifications 282...
Page 312: ...14 4 Disposing of the Instrument 304...
Page 349: ...Appendix 18 Outline Drawing A37 Appendix Appendix 18 Outline Drawing...
Page 362: ...Index Index 4...
Page 363: ......
Page 364: ......