Q1. The load voltage indicated on the load module is below expected.
A1. Ensure the load leads are as short as possible, twisted and use
the appropriate wire gauge. Ensure that voltage sense is used, this
can help alleviate the voltage drop across the load the leads.
Q2. When I try to start a program sequence, it will not run. “No Active
Channel” is displayed.
A2. Ensure the channel(s) is activated (not set to OFF) in the
FUNC
Program
Active Channel menu.
Q3. When trying to save to USB, the USB memory stick is unresponsive.
A3. Try restarting the PEL-2000B mainframe. If this fails to solve
the problem, ensure the USB memory is cleanly formatted.
Q4. When I try to clear an alarm, it doesn’t work.
A4. Before clearing an alarm or using the Protection Clear All
function, the DUT must be turned off. After the DUT is off, the
alarm(s) can be cleared.