Changing the operating mode will only affect the
current (active) channel. Other channels will not be
affected by any changes.
Editing CP Parameters
Background
Constant Power mode can be set to have a
maximum limit (Curr Limit). Using the current
limit enables limiting the current draw.
When using CP mode on a single channel load
module, two power levels can be configured, A
Value and B Value. On a dual channel load
module, only one power level can be configured
per channel: A Value.
Ensure the menu is in CP Mode. See page 138.
Configure
Configure
Mode
CP
Range
Low
Parameters
A Value
0 ~ Setting Max W
B Value
0 ~ Setting Max W
Curr Limit Load module dependent
Panel Operation
1.
Use the Selector knob to highlight
A Value.