PEL-2000B Series User Manual
134
Parameters
Level1
Minimum ~ Rating Ω
Level2
Minimum ~ Rating Ω
SlewRate
Load module dependent
SlewRate
Timer1
0.025 ~ 30000.0ms
Timer2
0.025 ~ 30000.0ms
Note
When used in Group Unit Mode, the Level1 & Level2
range is the combined rating of all the units used in
Group Unit Mode.
Panel Operation
1.
Use the Selector knob to highlight
Level1.
CRDL
Configure
Configure
Mode
CR
Dynamic
CH1
Range
Low
Level2
SlewRate
SlewRate
Timer1
Timer2
Level1
100.000
100.000
0.40
0.40
0.025
0.025
Ω
Ω
A/uS
A/uS
mS
mS
main
80V
05/04/22
16 : 50
2.
Press the Selector knob to edit the
selected level, then turn to
increase or decrease the value*.
OR
Use the number pad to enter a
number.
9
P9
6
P6
P3
CLEAR
LOCK
3
8
P8
5
P5
P2
CAL.
2
7
P7
4
P4
P1
0
P0
1